DODParts.com distributes related products varnishes dopes paints, aircraft parts and consumables, electrical parts, fittings and other components for the aerospace and defense industries. Submit the short price quote form for NSN 6632-01-332-7885 to receive an up to date pricing and availability quote by email.
A microprocessor based instrument incorporating more than one method of analysis such as colorimetric and potentiometric/ion selective or kinetic enzyme. It may be either a stand-alone unit or modular in design. For instruments which have only one method of analysis see analyzer, as modified. Excludes:all instruments which employ a chromatographic method of analysis.
50.0 hertz nominal or 60.0 hertz nominal frequency rating, blood gases and electrolytes; hco3; hct; base excess and o2 sat test type for which designed, 8.400 inches height, ac current type, digital display screen and integral printer and self-calibration features provided
Product Characteristics
Below you’ll find the complete technical specifications and performance data for this part, including material details and compliance information.
- Height
- 8.400 inches
- Current Type
- ac
- Frequency Rating
- 50.0 hertz nominal or 60.0 hertz nominal
- Test Type For Which Designed
- blood gases and electrolytes; hco3; hct; base excess and o2 sat
- Features Provided
- digital display screen and integral printer and self-calibration
- Part Name Assigned By Controlling Agency
- blood gas analyzer 120/230v 50/60 hz ac:
- Method Of Analysis
- polarographic and potentiometric/ion selective electrode
- Analysis Test/Measurement Range
- po2 0-600 mmhg; ph 6.8 to 7.8; pco2 8 to 99 mmhg; k 0.1 to 9.9 mm/l; na 100 to 200 mm/l; ca 0.25 to 4.25 mm/l; hct 10 to 75 pct; hco3 3.0 to 60 mm/l; tco2 3.0 to 60 mm/l; base excess -30 to +30 mm/l; o2 sat 30 to 99 pct
- Test Sample/Specimen Size
- 0.5 ml
- Test Sample/Specimen Type
- whole blood
- Depth
- 18.300 inches
- Voltage In Volts
- 120.0 and 230.0
- Special Features
- utilizes a disposable cartridge/test pack that contains all electrodes, calibrating and flush solutions and gases plus a waste container for processed blood, calibrating, and qc solutions. the cartridge/test pack shall eliminate the need for gas tanks, external solutions, and electrode maintenance. shall be pre-programed to automatically perform two-point calibration every hour and one-point calibration every 90 seconds. capable of automatic aspiration of the sample,and to process, display, and print the results in less than 2 minutes. shall be suitable for the field environment.
- Width
- 9.100 inches
Related Parts
This NSN group includes multiple components designed to meet specific performance and operational standards. Each listing provides detailed specs, functionality data, and replacement compatibility information:
| Part Number | Name |
|---|---|
| 90-0978P | Blood Gas Analyzer |
| 93-1046 (IRR) | Blood Gas Analyzer |
FAQs
Find helpful answers about installation, operation, maintenance, compliance, and troubleshooting.
What is the FSCG of the NSN 6632-01-332-7885?
Does NSN 6632-01-332-7885 require demilitarization?
Does NSN 6632-01-332-7885 contain precious materials?
Does NSN 6632-01-332-7885 contain hazardous materials?
When the NSN 6632-01-332-7885 was assigned to the Federal Catalog?
Professional Distribution
Certified to meet the strict requirements of defense, aerospace, and hazardous materials distribution.
- ITAR Registered
- DDTC Registered
- HAZMAT Certified
- AS9120 Quality Standards
- 100% Quality Inspections
You May Also Like
Browse other in-demand NSN parts sourced by professionals in defense, aerospace, and manufacturing sectors. Each component meets strict standards for quality and performance.
Pricing
Submit your request below and our team will respond quickly with up-to-date pricing, delivery estimates, and stock information: